SECTION 9—DIAGNOSTIC CODES
NO. OF
DIAGNOSTICS
BGD BARS
CODE
FLASHING
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
*NOTE: Refer to MK5 TT-EX User Advices on page 39 for more information.
™
™
™
MK
EX
and MK
TT-EX
5
5
MK5 TT-EX DIAGNOSTIC CODES
USER
ADVICE*
E 40
0
E 41
1
E 42
2
E 43
2
E 44
2
E 45
2
E 46
5
E 47
5
E 48
2
E 49
2
E 50
2
E 51
2
E 52
5
E 53
5
E 54
5
E 55
5
E 56
5
E 57
5
E 58
5
E 59
5
E 60
5
E 61
5
E 62
6
E 63
6
E 64
6
E 65
6
E 66
3
E67
4
E 68
8
E 69
9
E 70
8
E 71
9
E 72
7
E 73
7
E 74
7
E 75
7
E 76
7
E 77
7
E 78
11
™
Electronics
CAUSE OF ERROR
Unidentifiable error
Replace control only if fault repeats.
M2 Current Sensor Error
M2 Current Sensor Error
M1 Current Sensor Error
M1 Current Sensor Error
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
M2 Current Sensor Error
M2 Current Sensor Error
M1 Current Sensor Error
M1 Current Sensor Error
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
Current Calibration Error (Factory Only)
M2 Motor Over Current fault
M1 Motor Over Current fault
M2 motor - too much drag/load
M1 motor - too much drag/load
M2 motor - too much drag/load
M1 motor - too much drag/load
Replace control - Software error
Replace control - Software error
Replace control - Software error
Replace control - Software error
Replace control - Software error
Replace control - Software error
Internal power supply fault - replace control
40
Part No 1114808