Replacing The Fuses - UNI-T UT60B Operating Manual

Hide thumbs Also See for UT60B:
Table of Contents

Advertisement

D. Replacing the Fuses (see figure 13)
(figure 13)
Warning
To avoid electrical shock or arc blast, or personal injury or
damage to the Meter, use specified fuses ONLY in
accordance with the following procedure.
To replace the Meter's fuse:
1.
Press the POWER to turn the Meter off and remove all
connections from the terminals.
Remove the screw from the battery compartment, and
2.
separate the battery compartment from the case bottom.
3.
Remove the 2 rubber feet and 2 screws from the case bottom,
and separate the case top from the case bottom.
4.
Remove the fuse by gently prying one end loose, then take
out the fuse from its bracket.
5.
Install ONLY replacement fuses with the identical type and
specification as follows and make sure the fuse is fixed
firmly in the bracket.
Fuse 1: Glass fuse 0.5A, 250V, fast type, 5x20mm.
Fuse 2: Glass fuse 10A, 250V, fast type, 5x20mm.
Model UT60B/C/E: OPERATING MANUAL
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ut60cUt60e

Table of Contents