Appendix C Using the Digital Port

Fault/Inhibit Operating Mode

202
From the remote interface:
To configure the Inhibit In function for pin 3:
DIG:PIN3:FUNC INH
To configure the pin polarity to positive for pin 3:
DIG:PIN3:POL POS
After you have configured pin 3 as the remote inhibit input, you must
also configure the operating mode of the inhibit signal.
From the front panel:
Press the Settings key to access the Source Settings window.
Navigate to and select Protection, then Advanced. Then press Enter.
Select the Inhibit dropdown list. The inhibit signal can be live,
latched, or off. The Inhibit mode is stored in non-volatile memory.
Inhibit Mode
Description
Live
Allows the enabled outputs to follow the state of the Inhibit
input. When the Inhibit input is true, the outputs are disabled.
When the Inhibit input is false, the outputs are re-enabled.
Latched
Causes a logic-true transition on the Inhibit input to disable all
outputs, which will remain disabled.
Off
The Inhibit input is ignored.
From the remote interface:
To latch the Inhibit signal, send:
OUTP:INH:MODE LATC
To set the Inhibit signal live, send:
OUTP:INH:MODE LIVE
To disable the Inhibit signal, send:
OUTP:INH:MODE OFF
Model N6705 User's Guide