Performance Verification Testing

Required Test Equipment

Description
Probe Power Supply
Digital Multimeter
(DMM)
Signal Generator
Adapter
Test Coil
N2780/81/82/83 User's and Service Guide
Critical Specifications
No substitute
AC/DC voltage and current measurement
accuracy better than ±0.1% of reading at 10 mA
Input resistance in AC/DC voltage mode ¡ 1 M
DC to 66 Hz sine waves able to generate more
than 10 A in the test coil (e.g. an inductive load).
BNC (f) to dual stacking banana plug
5, 50, or 77 turns
(See coil construction instructions that follow.)
Test Coil Construction for the N2780A/B
Construct a 77- turn coil with a diameter of about 4 inches
(10 cm) using transformer wire.
Test Coil Construction for the N2781A/B
Construct a 50- turn coil with a diameter of about 4 inches
(10 cm) using transformer wire.
Recommended
Agilent N2779A
Agilent 34401A
Agilent 6813B
Pomona 1269
16-AWG 155C CU round
magnetic wire (NEMA MW
35-C); Red enamel
(transformer wire)
77 turns
50 turns