Introductory Programs

AFG Self-Test

46
Getting Started
The introductory programs in this section include:

AFG Self-Test

Resetting the AFG and clearing its status registers
Querying the AFG power-on/reset settings
Checking for Errors
Generating a sine wave with a single command
The AFG self-test is executed with the command:
*TST?
The AFG parameters tested include:
internal interrupt lines
waveform select RAM
segment sequence RAM
waveform segment RAM
DDS/NCO operation
sine wave generation
arbitrary waveform generation
marker generation
waveform cycle and arm counters
sweep timer
frequency-shift keying
stop trigger
DC analog parameters (amplitude, offset, attenuators, filters,
calibration DACs)
Upon completion of the test, one of the self-test codes listed in Table 1-2 is
returned.
Table 1-2. Agilent E1445A Self-Test Codes
Self-Test Code
0
Test passed
1
Test failed. An error message describes the failure.