Performance Test

Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)
Performance Test
This section does not establish preset SWR test procedures since there are
several test methods and different equipment available for testing the SWR
or reflection coefficient. Therefore, the actual accuracy of the test
equipment must be accounted for when measuring against instrument
specifications to determine a pass or fail condition. The test system used
must not exceed the system Rho uncertainties shown in the following tables
when testing the Agilent E- series E9300 power sensors.
Caution
DC voltages in excess of the maximum value (5 Vdc) can damage the sensing diode.
Table 4-1 Power Sensor SWR and Reflection Coefficient for the E9300A
Frequency
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
Table 4-2 Power Sensor SWR and Reflection Coefficient for the E9301A
Frequency
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 6 GHz
System Rho
Actual
Uncertainty
Measurement
±0.010
±0.010
±0.010
±0.010
±0.010
System Rho
Actual
Uncertainty
Measurement
±0.010
±0.010
±0.010
Service
Performance Test
Maximum
Rho
0.070
0.061
0.087
0.099
0.115
Maximum
Rho
0.070
0.061
0.087
59