Test Equipment - Agilent Technologies N1921A Operating And Service Manual

P-series wideband power sensor
Hide thumbs Also See for N1921A:
Table of Contents

Advertisement

Instrument
Signal generator
(Rise/fall time)
Signal Generator
(Zero set)
Oscilloscope
Network analyzer
Calibration kit
Diode detector
(Negative)
Power splitter
Power meter
Power sensor
Wideband power sensor
Cable adapter

Test Equipment

The following equipment are required for the performance tests:
Critical specification
Power range: –50 dBm to +22 dBm at 1 GHz
Output resistance = 50 Ω
Power range: –50 dBm to +20 dBm at 1 GHz
Output resistance = 50 Ω
Analog Bandwidth: 1 GHz and above
Frequency Range: 10 MHz to 40 GHz or
above 94 dB of dynamic range
<0.006 dB trace noise
Frequency Range: DC to 18 GHz or above
SMB connector
Two- Resistor type Power Splitter, N-type (f)
Max Frequency: 18 GHz
Dual Channel Peak Power Meter
Compatible with P Series and N8480 Series
Power Sensor
Absolute Accuracy: ±0.8%
Frequency: 50 MHz or above
Power range: –30 dBm to +20 dBm
SWR: ≤1.15 at 50 MHz
Power range: –30 dBm to +20 dBm
SWR: ≤1.20 at 1 GHz
11730A sensor cable adapter
12
Recommended Agilent model
number/part number
N5182A/E8257D/E8267D
E4438C/E8257D/E8267D
Infiniium MSO 9104A
E8361A/E8361C/E8363B/
E8363C
85054A/85054D/85056A/
85056D
33334EZ
11667A
N1912A
N8481A/N8487A
N1921A/N1922A
N1911A-200