Teledyne PEM 9002 Instruction Manual page 14

Portable emissions onitor
Table of Contents

Advertisement

Options
Nitrogen monoxide (NO) Measurement
Range:
Accuracy:
Resolution:
Sensor:
Response time (T90):
CO Measurement (without
Range:
Resolution:
Sensor:
Response time (T90):
Sulphur dioxide (S
O2) Meas
Range:
Accuracy:
Resolution:
Sensor:
Response time (T90):
0 to 2.000 ppm
5 ppm (up to 150 ppm)
5 % of reading (up to 2.000 ppm)
1 ppm
Electro-chemical cell
< 60 sec
H2
compensation)
0 ... 1.0 % Volume (10.000 ppm)
0.01 % Volume
Electro-chemical cell
< 60 sec
urement
0 to 2.000 ppm
5 ppm (up to 150 ppm)
5 % of reading (up to 2.000 ppm)
1 ppm
Electro-chemical cell
< 60 sec
9