1-3 Specifications

Performance specifications
General
Power supply
Power consumption
Protection against electrical shock
Display
EMC
Blood pressure measurement
Blood pressure measuring method
Pressure detection method
Pressure range
Accuracy
Measurement range
Pressurizing method
Air pressure control method
Rapid air exhaust system
Safety mechanism
Communications functions
Environment specifications
Operating temperature and humidity
Storage/transporting temperature and humidity
Physical specifications
Overall dimensions
Weight
Table 1 Specifications
120VAC 60 Hz
230VAC 50 Hz
(factory-preset)
Approx. 40VA
Class I , Type B
LED
EN 60601-1-2
Oscillometric method
Capacitance type pressure transducer
0 – 300 mmHg
Pressure: ±3 mmHg
Pulse rate: ±5%
Blood pressure: 10 – 280 mmHg
Pulse rate: 30 – 200 bpm
Micro pump
Rubber valve, ceramic valve
Electromagnetic valve
The electromagnetic valve is released when
approx. 320 mmHg or greater is detected.
Serial output: RS232C level
10 to 40°C, 85%RH or less, non condensing
-20 to 60°C, 95%RH or less, non condensing
245(W) X 322(H) X 390(D) mm
Approx. 9 kg