9-2 Replacing the Fuses

Replace the fuses as follows:
1. Remove the caps of both fuse holders located on the rear
panel.
2. Replace both fuses with new ones.
3. Re-install the fuse holder caps.
Use only fuses of the proper type, voltage and current rating as specified on the rear panel, and that
conform to IEC60127.

9-3 Checking the Counter

The TM-2655/TM-2655P is equipped with a counter function, which indicates how many times the device
has measured blood pressure. The counter data remains in memory even after the power is turned off.
Use the counter function to check the frequency of the device use or as the guideline for a periodic
cleaning

Displaying the counter

Press the COUNTER switch. The counter data appears in the
SYSTOLIC and DIASTOLIC displays for about 5 seconds.
Do not keep the COUNTER switch pressed. It will reset the counter
data to zero.
Ex. The measurement count at the figure is 2382 times.
Note
Note
The number of more than 1000 is displayed.
The number of less than 1000 is displayed.
The symbol is displayed at checking the counter.
22
Fig.19 Fuse holders