Specifications; Recorder - A&D TM-2440 Instruction Manual

Recorder for ambulatory blood pressure monitor
Hide thumbs Also See for TM-2440:
Table of Contents

Advertisement

4. Specifications

4.1. Recorder

Items
Measurement method
Pressure detection
method
Pressure display range
Measurement accuracy
Minimum
display division
Measurement range
Depressurization
Exhaust
Pressurization method
Automatic pressurization 85 to 299 mmHg
Interval time ( of A-BPM )
Clock
Display
Measurement count
Memory
Oscillometric measurement method
Semiconductor pressure sensor
0 to 299 mmHg
Pressure :
Pulse rate :
Pressure :
Pulse rate :
Systolic pressure : 60 to 280 mmHg
Diastolic pressure : 30 to 160 mmHg
Pulse rate :
Constant exhaust with controlled leakage valve
for safety mechanism
Electromagnetic valve
Micro-pump
Intervals at each section which divides 24
hours to six parts at the maximum.
Interval : OFF, 5, 10, 15, 20, 30, 60, 120 minutes
24 hour clock
OLED, 96 x 39 pixels, white characters
200 times or more. It varies due to
measurement conditions.
Measurement data : 600 data max.
11
Descriptions
(299 mmHg or more is hidden )
±3 mmHg
±5 %
1 mmHg
1 beat / minute
30 to 200 beat / minute