Teledyne GFC7000EU Manual page 4

Co2 analyzer
Table of Contents

Advertisement

MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE
(Addendum to GFC7000E Manual PN 04584)
There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system
okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2
exceeded. The relay alarm output AL1 is enabled all the time, whereas the AL2 and
AL3 can be enabled/disabled by going into the diagnostics menu (with 929
password), then factory options, then turning conc. Alarm Relays ON or OFF. When
the concentration alarm relays are enabled, concentration alarms status is no longer
available through the "Status output" pins. So, either the alarm relays or the status bits
could be used to monitor the concentration alarms but not both at the same time. The
default factory setting is to enable the concentration alarm relays. The AL1 relay is
energized when the system is okay and de-energized when the system has a fault. The
AL2 and AL3 relays energize when the corresponding concentration limits are
exceeded.
11.1.1. Interpreting WARNING Messages
The most common and/or serious instrument failures will result in a warning message
being displayed on the front panel. The table below lists warning messages, along with
their meaning and recommended corrective action.
It should be noted that if more than two or three warning messages occur at the same
time, it is often an indication that some fundamental analyzer sub-system (power supply,
Relay Board, Motherboard) has failed rather than indication of the of the specific failures
referenced by the warnings. In this case, it is recommended that proper operation of
power supplies, the Relay Board, and the A/D Board be confirmed before addressing the
specific warning messages.
The analyzer will alert the user that a Warning Message is active by displaying the
keypad label MSG on the Front Panel. In this case the Front panel display will look
something like the following:
SAMPLE
RANGE=50.00 PPM
<TST TST> CAL
The analyzer will also alert the user via the Serial I/O COM port(s) and cause the FAULT
LED on the Front panel to blink.
To View or Clear the various warning messages press:
CO = 00.00
MSG
CLR SETUP
4
05134 Rev -

Advertisement

Table of Contents
loading

Table of Contents