Agilent Technologies U1242A User's Manual And Service Manual page 68

Handheld digital multimeters
Hide thumbs Also See for U1242A:
Table of Contents

Advertisement

5
Performance Tests and Calibration
Step
Test Function
6
Turn the rotary switch to
μA
position
7
Press
to go to
function
8
Turn the rotary switch to
mA
position
9
Press
to go to
function
54
Cal Range
Input
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
1000 μA
1000 μA
10000 μA
10000 μA
μA
1000 μA
50 μA, 70 Hz
100 μA, 70 Hz
1000 μA, 70 Hz
10000 μA
1000 μA, 70 Hz
10000 μA, 70 Hz
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
100 mA
100 mA
1000 mA
320 mA
mA
100 mA
5 mA, 70 Hz
10 mA, 70 Hz
100 mA, 70 Hz
1000 mA
100 mA, 70 Hz
320 mA, 70 Hz
U1241A
U1241A/U1242A User's and Service Guide
Cal Item
U1242A
oPEn
1000.0 μA
10000 μA
50.0 μA
100.0 μA
1000.0 μA
1000 μA
10000 μA
oPEn
100.00 mA
320.0 mA
5.00 mA
10.00 mA
100.00 mA
100.0 mA
320.0 mA

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1241a

Table of Contents