4
Service and Maintenance
3 Gently remove the defective Fuse 1 by prying one end of the fuse and
4 Replace a new fuse of the same size and rating into the center of the
Figure 4-10 Fuse 1 replacement
5 If you want to replace a defective Fuse 2, remove Fuse 1 prior to
6 Gently remove the defective Fuse 2 by prying one end of the fuse loose
7 Replace a new fuse of the same size and rating into the center of the
Figure 4-11 Fuse 2 replacement
32
removing it out of the fuse bracket, see
fuse holder.
removing the Fuse 2. Loosen the four screws (shown in
lift and remove the circuit board from top case.
and removing it out of the fuse bracket, see
fuse holder.
Fuse 1 (440 mA/1000V)
Fuse 2 (11 A/1000 V)
U1241A/U1242A User's and Service Guide
Figure 4-
10.
Figure 4-
Figure 4-
11.
Rotary Switch
11) to