Agilent Technologies U1242A User's Manual And Service Manual page 58

Handheld digital multimeters
Hide thumbs Also See for U1242A:
Table of Contents

Advertisement

5
Performance Tests and Calibration
Step
Test Function
5
Turn the rotary switch to
6
Turn the rotary switch to
7
Turn the rotary switch to
position
8
Press
to go to
function
9
Turn the rotary switch to
position
10
Press
to go to
function
44
Range
1000 W
position
10 kW
100 kW
1000 kW
10 MW
100 MW
position
1000 nF
10 μF
100 μF
1000 μF
10 mF
μA
1000 μA
10000 μA
μA
1000 μA
10000 μA
mA
100 mA
440 mA
mA
100 mA
440 mA
Error from nominal 1 year
5520A Output
U1241A
1000.0 W
10.000 kW
100.00 kW
1000.0 kW
10.000 MW
100.00 MW
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
1000.0 μA
10000 μA
1000.0 μA, 500 Hz
1000.0 μA, 1 kHz
10000 μA, 500 Hz
10000 μA, 1 kHz
100.0 mA
[3]
400.0 mA
100.00 mA, 500 Hz
100.00 mA, 1 kHz
[3]
400 mA
, 500 Hz
[3]
400 mA
, 1 kHz
U1241A/U1242A User's and Service Guide
U1242A
[1]
± 3.3 W
[1]
± 33 W
± 330 W
± 3.3 kW
± 83 kW
[2]
± 1.53 MW
± 12.4 nF
± 0.124 μF
± 1.24 μF
± 20.4 μF
0.204 mF
± 1.3 μA
± 13 μA
± 10.5 μA
± 15.5 μA
± 105 μA
± 155 μA
± 0.23 mA
± 2.3 mA
± 1.05 mA
± 1.55 mA
± 4.5 mA
± 6.5 mA