Agilent Technologies U1242A User's Manual And Service Manual page 67

Handheld digital multimeters
Hide thumbs Also See for U1242A:
Table of Contents

Advertisement

Step
Test Function
3
Turn the rotary switch to
position
4
Turn the rotary switch to
position
5
Turn the rotary switch to
position
U1241A/U1242A User's and Service Guide
Cal Range
Input
Short
Dual banana plug with
copper wires short
between two terminals
1 V
1 V
Short
Dual banana plug with
copper wires short
between two terminals
10 MW
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
10 MW
1000 kW
1000 kW
100 kW
100 kW
10 kW
10 kW
1000 W
1000 W
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
1000 nF
400 nF
1000 nF
10 μF
10 μF
100 μF
100 μF
1000 μF
1000 μF
10 mF
10 mF
Performance Tests and Calibration
Cal Item
U1241A
1.000 V
10.000 MW
1000.0 kW
100.00 kW
10.000 kW
1000 W
400.0 nF
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
5
U1242A
SHrt
SHrt
oPEn
oPEn
53

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1241a

Table of Contents