Agilent Technologies U1242A User's Manual And Service Manual page 29

Handheld digital multimeters
Hide thumbs Also See for U1242A:
Table of Contents

Advertisement

Press
at
T1T2 (U1242A)
position to
enable T2
function
If you are working in a varied environment, where ambient temperature is
not constant, do the following:
1 Press
2 Avoid contact between the thermocouple probe and the measurement
3 After a constant reading is obtained, press
4 Touch the measurement surface with the thermocouple probe.
5 Read the display for the relative temperature.
The T2 temperature measurement is only available for U1242A.
N O T E
U1241A/U1242A User's and Service Guide
for more than one second to enable 0 °C compensation.
This function allows a quick measurement of the relative temperature.
surface.
the relative reference temperature.
Features and Functions 2
Measuring tips:
• Clean the measurement surface
and remember to disable the
applied power.
• When measuring temperature, move
the thermocouple along the surface
until you get the highest/lowest
temperature reading.
• For quick measurement, use the
0 °C compensation to see the
temperature variation of the
thermocouple sensor. The 0 °C
compensation assists you in
measuring relative temperature.
to set the reading as
15