Datamax W-6208 Operator's Manual page 70

W class series
Hide thumbs Also See for W-6208:
Table of Contents

Advertisement

Advanced Entry Calibration (continued)
Step
Operator Action
6
Press the
to scroll to 'Advanced
Entry'.
7
Press the
8
Press the
9
Press the
Place the label under the
Sensor Eye Mark and
lower the printhead
assembly.
10 Press the
to select the setting
(denoted by an '*') and
then record the sensor
reading (the 'yyy' value)
in a table similar to the
one shown after Step 12,
as a Label Value.
Next, press the
key to increment the
Gain Number.
Repeat this step for each
of the Gain Numbers
(00-32).
60
Displayed Message
key
key.
key.
key.
key
-OR-
-OR-
Comment
You are entering
Advanced Entry
Calibration.
Press the
key to abort this
procedure.
You are beginning
the Advanced
Entry Calibration.
If using preprinted
media, ensure the
label area under
the sensor is free
of preprinted text,
graphics or
borders.
Where 'yyy' is a
numerical value
representing the
current sensor
reading: useful in
locating the best
sensor position.
Note: Never
position the
sensor over a
perforation when
recording a
sensor reading.