Materials Required

Calibration and
Frequency Ranges
Table 4-1
Performance Verification
The following materials are required to run the tests:
8510XF system with accessories
85059A 1.0 mm Precision Calibration and Verification Kit
PC-based, 8510 Specification and Performance Verification Software
(Rev. A.05.00 or greater; p/n 08510-10033)
Laptop or PC running BASIC for Windows (3.1/95/NT)
(Rev. 6.3 or greater)
GPIB Card for PCs (National Instruments or HP)
PCMCIA Card for Omnibook or laptop
GPIB cable from PC to 8510XF
The 8510XF requires separate verifications for frequencies below and above
50 GHz (see on
"Frequency ranges" on page
A special two-band calibration procedure is required for the performance
verification of the 8510XF system (a measurement calibration is not
sufficient). There are up to three different calibrations and corresponding
frequency ranges required, which are stored in suggested locations in the
8510 analyzer (Refer to
Table
on
"Frequency ranges" on page
are set up automatically on the analyzer by the performance verification
software.
Calibration storage locations
File Location Name:
Cal Set 1
Cal Set 2
Cal Set 3
1. This calibration is for Option 005 systems only.
2. The frequency for standard systems is 2-50 GHz.
3. The frequency range shown is for the E7352A system only. The E7342A
system's frequency range will be: 50-85 GHz.
4-2 8510XF Service Quick Reference Guide
4-3).
4-1). For more information refer to the section
4-3. The parameters for these calibrations
Frequency Range of Data
Stored:
1
2
1-50 GHz
(2-50 GHz)
1
45 MHz
3
50-110 GHz