Hardware States; Instrument States - Agilent Technologies 8510XF Service Quick Reference Manual

Network analyzer system
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Sweep Time

Hardware States

Table 5-3

Instrument States

Table 5-4
System Parameter
RF Power source #1
LO Power source #2
Z
o
Delay
1. The 8510XF normally operates in step mode (ramp mode is not available).
2. Source power is set/read with internal source leveling selected. Normal (default) operation of the 8510XF is with
3. Z
Similar to 8510C.
The hardware states specific to the 8510XF are as shown in
Hardware States Specific to 8510XF
System Parameter
1
System Phaselock
1
Multi Source
Leveling Source #1 (RF)
2
Leveling Source #2
1. These menus are integrated into the firmware and are not available as softkeys.
2. Source leveling is set to system for normal (default) 8510XF operation. The source will indicate
external leveling for the RF source and internal leveling for the LO source.
Receiver: Constant frequency = 0.020000000 GHz
Table 5-4
shows the instrument states specific to the 8510XF system.
Instrument states specific to the 8510XF system
Frequency Range
(45 MHz to 50 GHz)
1,2
−10 dBm
1,2
+10 dBm
50 ohm
Coaxial
system source leveling selected.
is normally set to one Ohm for calibrations/measurements using waveguide test port connectors.
o
System Level Servicing & Troubleshooting
8510XF Firmware Specifications
Frequency Range
(45 MHz to 50 GHz)
None
On
2
System (External)
System (Internal)
Frequency Range
(50 to 85 GHz or
50 to 110 GHz)
−10 dBm
+10 dBm
3
50 ohm
Coaxial
8510XF Service Quick Reference Guide 5-7
Table
5-3.
Frequency Range
(50 to 85 GHz, or 50 to 110 GHz)
None
On
System (External)
System (Internal)