1. Introduction
1.1 General Information
This standard operating procedure (SOP) defines the procedure for the installation
qualification, operation qualification and performance verification (IQ/OQ/PV) of Agilent 3200
series meters. This procedure provides a guideline to generate the data necessary to validate
or qualify the Agilent 3200 series meters.
This SOP is applicable to all Agilent 3200 series meters that are used in a regulated lab. And
this SOP can be used as a reference for customer to develop their own IQ/OQ/PV documents.
It's customers' responsibility to develop and maintain the IQ/OQ/PV SOP for their specific
applications.
1.2 Meter Information
Installation qualification and operation qualification are required for all new meters used in a
regulated environment. The meters will not be released for regulated use until the installation
qualification and operation qualification have been performed with passing result.
Part Number
Description
G4383-64000
Agilent 3200P Benchtop pH Meter
G4384-64000
Agilent 3200C Benchtop Conductivity Meter
G4385-64000
Agilent 3200D Benchtop Dissolved Oxygen Meter
G4386-64000
Agilent 3200I Benchtop Ion Meter
G4387-64000
Agilent 3200M Benchtop Multi-Parameter Analyzer
1.3 Required Material
The following materials are required to perform the IQ/OQ/PV on Agilent 3200 series meters
Short circuit plug, PN: G4383-40000, shipped with 3200P/3200M/3200I
Conductivity diagnostic tool, PN:5185-8391, shipped with 3200C/3200M
ATC temperature diagnostic tool, PN: 5185-8390, shipped with
3200P/3200M/3200I/3200C/3200D
Manuals/User Guide
For 3200P/3200M with pH function:
Issued: 01-05-2012
Copyright 2012
Revision: 1.0
Page 3 of 13
Measurement Function
pH,ORP
Conductivity
Dissolved Oxygen
ISE
pH, ORP, Conductivity,
Dissolved Oxygen, ISE
Agilent Technologies