For measuring DC current
MEASURE
mA
FUSE
250V F 125mA
120mA MAX

Measuring the DC Voltage, DC Current, and Resistance

1
Press the [MEASURE ON] key.
2
Press the [FUNCTION] key on the MEASURE side to align
[DCA], or [Ω] that you want.
3
Press the [RANGE] key on the MEASURE side to select the measuring range.
Type
DCV
DCA
The measured result is displayed. The displayed value is updated every second.
Note
• Any data value being measured that exceeds the 120% of the measuring range will result in
overrange. The display then indicates [
range used).
• If no measured data are present immediately after the MEASURE key is turned on or if the
measurement function or range is switched to change the data on the display, the reading changes
to [
• When you switch the MEASURE key from the off to on, the instrument begins measurement with
the settings given immediately before the key was turned off.
• Providing any different type of setting with the [ZERO CLR] and [NEXT ENTER] keys during
measurement will abort measurement.
Display Holding On and Off
This determines whether the updating of displayed measured data is halted.
1
Press the [HOLD] key. "HOLD" is displayed.
2
To disable the display holding, press [HOLD] again. "HOLD" disappears from the
display.
Note
• The "HOLD" state only stops the updating of the display, but allows the instrument to continue
sampling data. This means that the measured data are updated through the communication
interface even during the "HOLD" state.
• Holding the display is not allowed if the communication mode is "talk-only" or the "printer"mode.
(See page 39.)
Hi
42V MAX
Measured
object
Lo
Measuring Range
Display
35 V
5 V
500 mV
100 mA
20 mA
50 kΩ
5 kΩ
500 Ω
].
with the [DCV],
V
V
mV
mA
mA
kΩ
kΩ
] (the position of the decimal point depends on the