RANGES, LINEARITY AND RESOLUTION
Sensor Type
and Material
Standard
Range (°C)
Linearity
error (°C)
Resolution
{°C)
J
lEC 584-1
-210 to-200
0.2
0.1
Fe-CuNi
(Fe-Const)
ASTM E230
-200 to 1200
0.1
0.1
K
lEC 584-1
-255 to -220
0.2
0.5
NiCr-NiAl
ASTM E230
-220 to -140
0.2
0.2
(Chromega™ -
Alomega™)
DIN 43710
-140 to 1372
0.1
0.1
T
lEC 584-1
-255 to -230
0.1
0.5
Cu-CuNi
ASTM E230
-230 to-160
0.1
0.2
(Cu-Const)
-160 to 400
0.1
0.1
E
lEC 584-1
-260 to -250
0.1
0.5
NiCr-CuNi
ASTM E230
-250 to -220
0.1
0.2
(Chromega-Const)
-220 to 1000
0.1
0.1
R
lEC 584-1
-50 to 250
0.1
0.5
R13%Rh-R
ASTM E230
250 to 1768
0.1
0.2
S
lEC 584-1
-50 to 600
0.1
Pt10%Rh-R
ASTM E230
DIN 43710
600 to 1760
0.1
B
lEC 584-1
400 to 1200
0.1
0.5
R30%Rh-R6%Rh
ASTM E230
1200 to 1820
0.1
0.2
N
Nicrosil-Nisil
NBS Mono¬
graph 161
Oto 1300
0.1
0.1
Fe-CuNi DIN
(Fe-Const)
DIN 43710
-200 to 900
0.2
0.1
Cu-CuNi DIN
(Cu-Cpnst)
DIN 43710
-200 to 600
0.1
R100
DIN 43760*
-200 to 200
200 to 1200
0.05
0.05
0.05
0.1
Range
Resolution
mV
-lOto llOmV
0.002 mV
ohm
0 to 400 ohms
0.01 ohm
*R100 range 850°C to laOC'C is extrapolated from DIN 43760 curve.