B) REFERENCE JUNCTION COMPENSATION USING A Pt100 PROBE
CJ MODE =PROBE
In this mode, the reference junction T2 is outside the CL521 at a location where a PttOO
probe is also installed. The thermocouple is connected to the CL521 with copper wires
and the PtIOO probe is connected to the PtIOO terminals.
I-'
Figure 4-9. Reference Junction Compensation Using a PtIOO Probe
C) MANUAL REFERENCE JUNCTION COMPENSATION, CJ MODE = MAN
Reference junction temperature compensation is automatically set to 0°C, using the
external reference junction. The thermoelectric potential is connected to the CL521
using copper leads.
Figure 4-10. Manual Reference Junction Compensation