Resistance Test; Diode Test - Fluke II Series Service Manual

Hide thumbs Also See for II Series:
Table of Contents

Advertisement

Maintenance
PERFORMANCE
TESTS
Resistance Test
3-10.
1.
Select the
ohms
function
on
the
UUT.
2.
Connect
the
Ohms
Calibrator
or
Decade
Resistor
to
the
vn-w- and
COM
input
terminals
of
the
UUT.
3.
Referring
to
Table
3-4, set
the
Ohms
Calibrator
or
Decade
Resistor
to the resistance
value
indicated
in steps
1
through
6.
Verify
that
the display
reading
is
within the
limits
shown.
Table
3-4.
Resistance Test
STEP
INPUT
RESISTANCE
DISPLAY
READING
1
short
0.0
to
0.2Q
Decades
of
1:
2
100Q
99.3
to
100.7Q
(plus
0
reading)
3
1
kQ
994
to
1006
Q
4
10
kQ
9.994
to
10.06
kQ
5
100
kQ
99.94
to
100.6
kQ
6
1
MQ
.994
to
1.006
MQ
7
10
MQ
9.79
to
10.21
MQ
Decades
of
1.9:
2
i9on
188.8
to
191.2Q
(plus
0
reading)
3
1.9
kQ
1889
to
1911
Q
4
19
kQ
18.89
to
19.11
kQ
5
190
kQ
188.9
to
191.1
kQ
6
1.9
MQ
1.889
to
1.911
MQ
7
19
MQ
18.61
to
19.39
MQ
Decades
of
2.7:
2
270Q
268.4
to
271.
6Q
(plus
0
reading)
3
2.7
kQ
2685
to
271
5Q
4
27
kQ
26.85
to
27.15
kQ
5
270
kQ
268.5
to
271.5
kQ
6
2.7
MQ
2.685
to
2.715
MQ
7
27
MQ
26.45
to
27.55
MQ
Diode Test
3-1
1
.
To
test
the Series
II
meters,
perform
the following
steps:
1.
Set
the
UUT
to
the
Diode
Test
function.
NOTE
On
Fluke
5100
series calibrators, activate the
50Q. divider
override.
2.
Apply an
input
from
the
DC
Voltage
Calibrator
of
+.090V
dc
to
the
vn-w- and
COM
input terminals
of
the
UUT,
and
verify
that
the
beeper
is
on.
3.
Increase the
DC
Voltage
Calibrator output
to +.1
10V
dc,
and
verify
that
the
beeper
is
off.
3-8