HP 1670A Series User Manual page 77

Logic analyzers
Table of Contents

Advertisement

Troubleshooting
Assess the problem
Problems transferring or copying files
Copying files out of the logic analyzer
• If
you have problems copying files out of the logic analyzer, you might be
experiencing timeout problems. See the timeout section on the previous
page.
• If
you only receive 1 byte back when copying files, refer to "Dynamic Files"
later in this chapter.
Copying files into the logic analyzer
• If
you have problems copying files into the logic analyzer, such as copying
setup or data to change a configuration, then check the File Timeout
setting in the LAN Settings menu. Refer to "Configure the network
addresses" in Chapter 1 for more information.
Communications not established
If
you have just installed and configured the LAN interface and you have
never been able to read the logic analyzer remote file server directory, go
directly to the troubleshooting section relating to your network in this
chapter.
If
you have been able to read the logic analyzer remote file server directory
and now cannot do so, check the following:
• Has any hardware been added or moved on your network? This includes
adding or removing any workstations or peripherals, or changing any
cabling.
• Have software applications been added to the network?
• Have any configuration files been modified?
• Have any of the following files been deleted or overwritten?
UNIX:
/etc/hosts
PCs:
I etc/inetd. conf
/etc/services
dependent network files
If
you know or suspect that something has changed on your network, check
the changes and adjust the configuration for the LAN interface using the
procedures in Chapter 1. Otherwise, proceed to the troubleshooting section
in this chapter relating to your network.