HP 1670A Series User Manual page 57

Logic analyzers
Table of Contents

Advertisement

Example
Programming the Logic Analyzer
To enter commands directly using telnet
Programming the logic analyzer over a telnet connection
To connect to the logic analyzer named 1660sys, enter:
$ telnet 1660sys 5025
The computer responds with:
Trying ...
Connected to 1660sys.col.hp.com.
Escape character is •A]'.
The connection was successful. Because the analyzer does not provide a
prompt, start entering programming commands. Typical commands might be:
:system:header on
:system:longform on
:select 1
:menu?
MENU 1,3
:system:dsp 'Triggering on memory violation'
:system:print screen
The small program above turns on the header and longform for query
responses, selects the analyzer, checks which menu it is on, creates a title for
that screen, and then prints it to the default printer.
When you are done, close the telnet connection. Enter the escape character
to get the telnet prompt. The escape character (Control and "]" in this
example) does not print.
telnet> quit
The telnet connection closes and you see your regular prompt.
Connection closed.
$