Figure 2-12: Limits Checking Application
Swept Source
OUT
Figure 2-13: Limits Checking Results
Amplitude
+10 dBm
o
o
o
+4 dBm
In this application a swept frequency signal is applied to the input of the
Device Under Test. The power meter measures the output power. The
limits have been set at +4 dBm and +10 dBm. A fail occurs each time the
output power is outside these limits as shown in Figure 2-12.

Setting Window Limits

The power meter can be configured to verify the current measurement in
either window against predefined upper and/or lower limit values. The
range of values that can be set for the upper and lower limits and the
HP E4419B User's Guide
Get other manuals https://www.bkmanuals.com
Device
Under Test
OUT
IN
Fail
o
o
o
o
Fail
Frequency
Power Meter Operation
Setting Measurement Limits
Power Meter
CHANNEL A
INPUT
2-47