Trane FARRAR ULC Series Installation, Operation And Maintenance Manual page 13

Ultra low storage chamber
Table of Contents

Advertisement

Construction
Chamber Exterior Material
Chamber Interior Material
Load Capacity per Chamber Side
Controller
Temperature Controller/QTY
Displays
Modbus TCP/IP
Alarms
Dry Contacts
Water-Cooler Condenser
Temperature Controller/QTY
Thermal Room Load: Water-Cooled
Air-Cooled Condenser
Painted Aluminum - White Embossed Finish
Painted Aluminum - White Embossed Finish
2,000 / 907 (4,000 / 1,814 total)
Microprocessor (2)
5.7 (in.) LCD (2)
Communication Protocol
Refrigeration system failure, high-low temperature, door ajar,
power failure, remote alarm contacts
Standard (C, NO, NC)
4 x 0.5 (in.) FPT connections (2 per refrigeration system
Pressure differential = approximately 20 PSI (1.4 bar)
Tower water:*
Water temperature 85°F. Maximum flow rate = 7 GPM (26.5 LPM)
Average flow rate = 4.5 GPM (17 LPM)
Chilled water:*
Water temperature ≤10°C
Maximum flow rate = 3.5 GPM (14 LPM)
Average flow rate = 2.0 GPM (8 LPM)
*Total for both refrigeration systems
Heat Rejection @ -80ºC Maximum:
= 28,000 Btu/hr. (8.2 kW/hr.) to water
= 1,800 Btu/hr. (0.52 kW/hr.) to ambient
Minimum Normal Operation, 1 system @ 75-80%:
= 12,000 Btu/hr. (3.5 kW/hr.) to water
> 900 Btu/hr. (0.26 kW/hr.) to ambient
Average - Normal Operation, 1-2 Door opening/hr.,1 System Defrost:
= 14,000 Btu/hr. (4.1 kW/hr.) to water
= 1,000 Btu/hr. (0.29 kW/hr.) to ambient
<1,550 Btu/hr.
0.44 kW/hr. Maximum
Heat Rejection @ -80ºC:
Maximum* = 28,000 Btu/hr. (8.2 kW/hr.)
Minimum = 12,000 Btu/hr. (3.5 kW/hr.)
Average = 14,000 Btu/hr. (4.1 kW/hr.)
*When both machine sections are running
ULC SERIES INSTRUCTION MANUAL
13