Storage Capacity and Dimensions (cont.)
ULC Series Model
Chamber Floor Space
sq. ft.
Door Opening
Chamber Wall Thickness
Ambient Conditions
Box Capacity
2 in. Freezer Boxes
3 in. Freezer Boxes
Vial Capacity
81 Vials in 2 in. box
81 Vials in 3 in. box
Inventory Racks for 2 in. boxes
Inventory Racks for 3 in. boxes
Application and Electrical Requirements
Application
Electrical ULC/ULCi
Electrical Requirements
North America
International
Circuit Breaker
Electrical Breaker
Temperature Control Range
Ambient Operating Temperature
Noise level
259
70.5
6.5
46 W x 67 H (in.) /
1,168 W x 1,702 H (mm)
x 2 doors
6 (in.)
152.4 (mm)
18°C to 30°C
70% RH Max
NOTE: FOR INDOOR
USE ONLY
Pollution Degree 2
3,920
2,520
317,250
204,120
112
(not included)
168
(not included)
Pharmaceutical, Biopharmaceutical, Biorepository, Bio-Logistics
208-230 VAC, 1 Ø, 60 Hz / 2
230 VAC, 1 Ø, 50 Hz / 2
Mains Supply Fluctuation +/- 10%
Overvoltage Category II
208-230VAC, 1 Ø, Single Phase, 60 Hz
230VAC, Single Phase, 50 Hz
2 x 30 amp
30A, 60 Hz. (32A, 50 Hz)/ 2 Circuits
-20°C to -80°C
+2°C to +8°C
+18°C to +30°C
65 dB(A)
ULC SERIES INSTRUCTION MANUAL
190
51.0
4.7
41 W x 87 H (in.) /
1,041 W x 2,210 H (mm)
x 2 doors
6 (in.)
152.4 (mm)
18°C to 30°C
70% RH Max
NOTE: FOR INDOOR
USE ONLY
Pollution Degree 2
3,430
2,058
277,830
166,698
70
98
311
67.8
6.3
44 W x 87 H (in.) /
1,118 W x 2,210 H (mm)
x 2 doors
6 (in.)
152.4 (mm)
18°C to 30°C
70% RH Max
NOTE: FOR INDOOR
USE ONLY
Pollution Degree 2
4,900
2,940
396,900
238,140
140
196
10