Trane FARRAR ULC Series Installation, Operation And Maintenance Manual page 11

Ultra low storage chamber
Table of Contents

Advertisement

ULC Series Material Handling Solution
ULC Model
Shelves
Type
Size (in.)
Size (mm)
Positioning
Product Cart
Qty/Type
Size (in.)
Size (mm)
Transfer Cart
Size (in.)
Size (mm)
Refrigeration Systems
Refrigeration Systems
Scroll Compressors
Blower Motor
Evaporator/QTY=2
Condenser/QTY=2
Metering Device
Refrigerant: High Stage
Refrigerant: Low Stage
Noise level
Requirements
259
Sliding (individually)
55 ⅛ x 40 ½
1,400 x 1,029
4 across by 6 high
@ 10 (in.) increments
6 shelves, 10 (in.) / 254 (mm)
49 ⅜ D x 20 ⅜ W x 51 ¼ H
1,229 x 517 x 1,302
41 ⅙ D x 39 W x 45 ⅛ H
1,043 D x 991 W x 1,146 H
ULC SERIES INSTRUCTION MANUAL
190
Stainless Steel (1 in.)
Weight Capacity:
300 lbs./ 136 kg.
54 x 36
1,372 x 914
Adjustable
@ 1 (in.) increments
4 shelves, 8 (in.) / 457 (mm)
36 ¾ D x 33 W x 61 ½ H
934 D x 838 W x 1,562 H
27 W x 56 D x 45 ⅛
686 W x 1,423 D x 1,146 H
Qty 2
Qty 4
Qty 2
500 cfm
14,158 l/min
Forced-Air
Forced-Air-Cooled or Water
Electronic
R449
R-290 & R508B (HFC)
65 dB(A)
*Tower water:
Water temperature ≤30°C
Maximum flow rate = 7 gpm (27 LPM)
Average flow rate = 4.5 gpm (17 LPM)
*Chilled water:
Water temperature ≤10°C
Maximum flow rate = 3.5 gpm (14 LOM)
Average flow rate = 2.0 gpm (8 LPM)
311
Pallet (capacity)
40 H x 48 W x 36 D
1,016 H x 1,219 W x 914 D
4 shelves, 18 (in.) / 457 (mm)
51 ½ D x 21 W x 61 ½ H
1,308 D x 534 W x 1,562 H
41
D x 39 W x 45 ⅛ H
1,043 D x 0991 W x 1,146 H
11

Advertisement

Table of Contents
loading