Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the physical-chemical characteristics of the substance according to the
Lambert-Beer Law:
Where:
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the
other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound
from a specific chemical reaction between sample and reagents. Given that the absorption of a
compound strictly depends on the wavelength of the incident light beam, a narrow spectral
bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's HI 96704 colorimeter is based on special subminiature LED and
narrow-band interference filters to guarantee both high performance and reliable results.
www.ponpe.com

PRINCIPLE OF OPERATION

-log I/I
-log I/I
=
Absorbance (A)
o
=
intensity of incident light beam
I
o
=
intensity of light beam after absorption
I
ε
molar extinction coefficient at wavelength λ
=
λ
c
=
molar concentration of the substance
d
=
optical path through the substance
HI 96704 block diagram (optical layout)
= ε
c d
λ
o
or
A = ε
c d
λ
6