Glp - Hanna Instruments HI 96704C Hydrazine ISM Instruction Manual

Hide thumbs Also See for HI 96704C Hydrazine ISM:
Table of Contents

Advertisement

• The instrument will return automatically
to measurement mode by displaying
dashes on the LCD.
In the GLP mode, the last user calibration date can be consulted and the factory calibration can
be restored.
LAST CALIBRATION DATE
To display the calibration date:
• Press GLP/
calibration month and day will appear on
the main display and the year on the
secondary display.
• If no calibration was performed, the
factory calibration message, "F.CAL" will
appear on the main display and the
instrument returns to measurement mode
after three seconds.
FACTORY CALIBRATION RESTORE
It is possible to delete the calibration and
restore factory calibration.
• Press GLP/ to enter GLP mode .
www.ponpe.com

GLP

to enter GLP mode . The
20