Acquisition (Data, STROBE, and PFI Channels)
Magnitude of differential input voltage .................0.05 V to 3.3 V
Positive-going differential threshold......................150 mV
Negative-going differential threshold ....................50 mV
Input impedance.....................................................100 Ω differential, nominal
Input range .............................................................–1.4 V to +3.8 V.
Power
Maximum power requirements when toggling half of the data channels at 100% data rate and half at
50% data rate. +12 V is maximized when generating, and +3.3 V is maximized when acquiring.
+12 V .....................................................................215 mA, 2.6 W
+3.3 V ....................................................................850 mA, 2.8 W
V
.......................................................................275 mA, 690 mW
ccoB
Physical
Dimensions ............................................................12.9 × 2.0 × 12.1 cm (5.1 × 0.8 × 4.7 in.)
Weight ....................................................................284 g (10 oz)
Front panel connectors...........................................One 68-pin VHDCI connector, and
Environmental
The NI 6583 is intended for indoor use only.
Operating environment ..........................................0 °C to 55 °C.
Relative humidity range .........................................10% to 90%, noncondensing.
Altitude ..................................................................2,000 m at 25 °C ambient temperature
Pollution Degree ....................................................2
Storage environment
Ambient temperature range ...........................–20 °C to 70 °C.
Relative humidity range .................................5% to 95%, noncondensing.
Note
Clean the device with a soft, non-metallic brush. Make sure that the device is completely dry
and free from contaminants before returning it to service.
NI 6583R User Guide and Specifications
one 73-pin InfiniBand connector
Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2.
Tested in accordance with IEC 60068-2-56.
Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2.
Tested in accordance with IEC 60068-2-56.