Turn on the power to the calibrator by toggling the switch on the power entry module. Aft
brief self-test, the co
ntroller should begin normal operation. The main screen will appear within 30
seconds. If the instrument fails to operate, please check the power connection.
The display will show the well temperature, and wait for user input before heating or cooling to
current set-point.
m Setup
Before using th
e instrument, the parameters in this section need to be setup (Main Menu:
SYSTEM|SETUP ME
NU).
y
Currently (2005), English is the only language supported by the instrument. Other languages are
plann
ed to be released in 2006. If other languages are present, select the preferred language using
the right/left arrow keys and press "ENTER" to accept the selection.
t
The contrast determines the readability and visibility of the text/numbers on the screen. Select the
de
sired contrast using the right/left arrow keys and press "ENTER" to accept the selection.
l
The decimal of the numbers in the instrument can be either a comma or decimal. Select the
d
esired decimal type using the right/left arrow keys and press "ENTER" to accept the selection.
ure
t
NOTE: Metrology Wells should not be expected to operate better than the stability limit
specification set forth in the Specifications section of this guide. Therefore, the minimum setting of
the stability limit should not be less than the stability specification set forth in the Specific
section.
The stability limit of the instrument is the parameter which allows the instrument to notify the user
when it h
as achieved the stability limit set in this parameter. There are two notifications: visual
and audible. The visual notification is always active. When the instrument is operating within the
stability limit, the stability parameter on the main screen will remain highlighted as long as the
instrument is within the given specification, otherwise the parameter will not be highlighted. The
audible, if enabled, alerts the user once per set-point when the instrument achieves the set stability