4
Remote release socket
without multi-format camera / without video printer
•
Connect the control cable at the MEMOSKOP "Printer" connector.
•
Cut off the 3.5 mm socket plug.
•
Remove the insulation from the ends of the wires.
•
Extend the cable using the 2-conductor cable included with the shipment.
•
Insulate the location of the new connections with shrink tubing.
•
Route the ends of the wires through the hole drilled for the remote release socket.
•
Solder the wires to the remote release socket (refer to 1/Fig.3 and 2/Fig.3).
•
Solder one end of the protective conductor included with the shipment to the remote
release socket (3/Fig.3) and connect the other end to the protective conductor terminal
screw (1/Fig.2).
•
Install the remote release socket.
When a multi-format camera is present
•
Cut off the 3.5 mm socket plug of the existing control cable.
•
Remove the insulation from the ends of the wires.
•
Extend the cable using the 2-conductor cable included with the shipment.
•
Insulate the location of the new connections using shrink tubing.
•
Route the ends of the wires through the hole drilled for the remote release socket.
•
Solder the wires to the remote release socket (refer to 1/Fig.3 and 2/Fig.3).
•
Install the remote release socket.
When a video printer is present
•
Cut off the remote control connector for the multi-format camera of the existing control
cable.
•
Remove the insulation from the ends of the wires (green and white wires).
•
Extend the green and white cable using the 2-conductor cable included with the
shipment.
•
Insulate the location of the new connections using shrink tubing.
•
Route the ends of the wires through the hole drilled for the remote release socket.
•
Solder the wires to the remote release socket (refer to 1/Fig.3 and 2/Fig.3).
•
Install the remote release socket.
SIREMOBIL Compact
Register 3
RXR2-130.033.02
Rev. 02
Connecting laser camera
Page 4 of 8
03.97
TD SP 1
1
1
1
1
Siemens AG
Medical Engineering