General
Safety Information
CAUTION
Tests or settings performed using radiation are indicated by the radiation symbol
Radiation protection clothing must be worn when performing such procedures.
Protective Conductor Measurement
After all work has been completed and all covers have been attached, perform a protec-
tive conductor measurement according to TI 236 / ARTD-002.731.17...
The protective conductor resistance must not exceed 0.2 Ohm.
Additional note
These instructions apply only to connecting a laser camera to a SIREMOBIL Compact.
Parts required
Laser camera connection kit
Measurement devices and auxiliary materials required
Service PC
Digital Multimeter
Service Software (from logbook)
Cable for connecting service PC to host
Protective conductor meter
Tools required
Standard tool kit (refer to ARTD / section 3)
1 set of Allen keys
Heat gun
Drill
17 mm drill bit
16 mm drill bit
3 mm drill bit
SIREMOBIL Compact
Ensure compliance with the product-specific safety information in the
documents as well as the general safety information found in Register 2
of the TI folder.
DMM
Register 3
RXR2-130.033.02
Rev. 02
Connecting laser camera
37 79 159 G5429
99 00 440 RE999
44 15 899 RV 090