Removing The Rear Cover - UNI-T UTS1000 Series Service Manual

Spectrum analyzers
Hide thumbs Also See for UTS1000 Series:
Table of Contents

Advertisement

Service Manual
3. For cleaning methods, see the "Maintenance" chapter.
Note: When the front panel is placed downward, it is necessary to avoid the knob cap to avoid
damage to the knob.

Removing the rear cover

The following procedure describes the removal and replacement of the rear cover.
prerequisites:
To prevent electrostatic damage to components, wear a properly grounded antistatic wrist and foot
strap during installation, and use an antistatic mat in a tested antistatic environment.
Steps:
1、 Remove all cables and power cords from the rear panel;
2、 Place the cushion flat on the electrostatic table;
3、 Place the instrument face down on a cushion to prevent damage to the screen and knobs;
4、 Pull up the handle, and use a T8 Torque screwdriver to remove the 2 screws on the case back cover
assembly, as shown in the figure below:
5、 Use a T8 torque screwdriver to remove the 8 screws on the case back cover assembly. As shown below:
6、 Carefully lift up the rear cover
7、 To reinstall, reverse the above steps.