Meter; Recorder - Fluke ScopeMeter 190 II Series 190-104 User Manual

Hide thumbs Also See for ScopeMeter 190 II Series 190-104:
Table of Contents

Advertisement

Fluke 190-104, Fluke190-204
Users Manual
Temperature (TEMP)
With Optional Temperature Probe (
Ranges (°C or °F) .................................. -40.0 to +100.0 °
Probe Sensitivity ............................ 1 mV/°C and 1 mV/°F
Accuracy........................................... ±(1.5 % + 5 counts)
(add temperature probe accuracy for overall accuracy)
Decibel (dB)
dBV................................................ dB relative to one volt
dBm ................... dB relative to one mW in 50 Ω or 600 Ω
dB on .......................................... VDC, VAC, or VAC+DC
Accuracy...........................same as VDC, VAC, VAC+DC

Meter

Four of the Automatic Scope Measurements as defined
above may be displayed at the same time, using larger
screen area for convenient reading, suppressing the scope
waveform information. For specifications see Automatic
scope Measurements above.
106
°
F not for Japan)
-100 to +250 °
-100 to +500 °
-100 to +1000 °
-100 to + 2500 °

Recorder

TrendPlot (Meter or Scope, Channel A ,B, C, D)
Chart recorder that plots a graph of min and max values of
Meter or Scope measurements over time.
Measurement Speed.........................>5 measurements/s
Time/Div............................................5 s/div to 30 min/div
Record Size (min, max, average)...............≥18000 points
Recorded Time Span ........................... 60 min to 22 days
Time Reference ..................... time from start, time of day
Scope Record
Records scope waveforms in deep memory while
displaying the waveform in Roll mode.
Source......................................................Input A, B, C, D
Max. Sample Speed (5 ms/div to 1 min/div) ...4x125 MS/s
Glitch capture (5 ms/div to 1 min/div) ......................... 8 ns
Glitch capture (2 min/div) ........................................... 8 ns
Time/Div in normal mode ................. 5 ms/div to 2 min/div
Record Size .........................30k to 37.5k points per trace
Recorded Time Span ................................ 6 s to 48 hours