Troubleshooting
Troubleshooting Assembly–Level Problems
Self
Tests 13xx: A11 Analog/Pulse Modulation Generator Self
Before proceeding to the reported self–test error code, check the following voltages. If
any of these voltages are out of specification troubleshoot the supply problem first.
Supply Voltage
Connector Pins
(Vdc)
+15
P132-28
P132-3
15
+5.2
P132-2, 27
P132-1, 26
5.2
+5.2 Digital high
P131-64, 65, 129, 130
+3.4 Digital Low
P131-60, 61, 62, 63,
125, 126, 127, 128
1300 Power Supply
• Replace the A7 Reference.
1-56
Minimum
Maximum
Value
(Vdc)
+14.55
+15.45
14.55
15.45
+5.04
+5.37
5.1
5.3
+5.04
+5.36
+3.29
+3.5
Test Errors
Origin
Value
(Vdc)
Main Supply
Main Supply
Main Supply
YIG Driver
Main Supply
Main Supply