Problem
The device alarms without
reason
Battery indicator doesn't
display when charging.
Battery icon flashes in
blue.
BW Icon
Shop for Gas products online at:
Cause
The sensor is exposed to a
puff of the target gas.
Alarm setpoints are set
incorrectly.
Sensors require calibration.
Missing or faulty sensors.
Battery temperature is out
of acceptable range.
Battery is depleted.
Battery is out of required
charging temperature
range.
CanarySense
www.
Apparatus is operating
normally. Use caution in
suspected areas. Check the
peak gas exposure reading.
Define the alarm setpoint in
Device Configurator.
Carry out calibration.
Replace the sensors.
Move to lower temperature
ambient to charge the battery.
Charge the battery for 8 hours.
If the battery indicator doesn't
light after charging, contact
Move to 0~45°C ambient
temperature.