4.
In your BW Icon:
Double press to enter the menu
l
Single press until the BLE icon is displayed
l
Hold press 3 secs to initiate the pairing mode.
l
5.
In the Device list screen from the Device Configurator app, select the BW Icon Serial
Number to start pairing.
6.
In your Mobile device, tap on the Menu button and then select Bump Test.
7. Enter the Operator Name, and then Tap Save. The IntelliFlash LED flashes amber to indicate
the bump test process has started.
8. In the Input Gas level screen, check the sensor that you want to test and enter the Span gas
concentration, and then tap START.
9. Open the cylinder valve by turning the pressure regulator knob counterclockwise. The Zero
process starts and a message is displayed when succeeded.
10. Follow onscreen instructions to know when to apply gas and when the bump test process is
complete.
11. The process is complete when the results are displayed on your mobile device. You can now
remove the cap by pulling on the tabs.