Honeywell BW Icon User Manual page 10

Portable multiple gas detector
Hide thumbs Also See for BW Icon:
Table of Contents

Advertisement

Move to a normal atmosphere (20.9% v/v O
test gas.
Details for calibration and maintenance:
Recommendations for calibration of the equipment on a routine basis including the
l
maximum time interval between calibrations.
Calibrate the apparatus before first-time use and then on a regular schedule, depending on
l
use and sensor exposure to poisons and contaminants. Honeywell recommends that the
sensors be calibrated regularly and at least once every 180 days (6 months).
The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different
l
combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.
Procedure to calibrate the detector via the calibration cap and the DC app on a
mobile device
1. Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs to
snap it into place. Wait a few minutes to sensors warm up.
BW Icon
Shop for Gas products online at:
) that is free of hazardous gas. Use 50% LEL for
2
CanarySense
www.
CAUTION
12
.com
User Manual
1.888.610.7664