Operation And Operating Modes Of Gaslab Q2; Operating Mode: Analysis (Automatic Normal Operation) - Honeywell enCore Series Manual

Gas quality analyzer
Hide thumbs Also See for enCore Series:
Table of Contents

Advertisement

7
7.4 Operation and operating modes of GasLab
Q2
The aim of this section is to explain more details of operation, operating modes and
device functions. First of all, the observation of the device functions and normal
operation are presented, which requires no or little action from the operating per-
sonnel. This is followed by operating instructions for more complex operating
modes and functions, for example, calibration.
7.4.1
Operating mode: Analysis (automatic normal
operation)
After connecting the power supply and reaching operating temperature, the
measurement will take place every second in an endless loop in the continuous
flow. The gas flow can be identified on the float flow meter. The Q2 main display will
be visible on the device and on the remote operation panel. "Analysis" will be output
as the operation and step on the display. Values will be displayed in black text. The
two LEDs will be permanently lit.
Apart from intervention by the operator, the analysis will only normally be in-
terrupted by the automatic operational calibration.
NOTICE AVIS
GasLab Q2 must perform an operational calibration at least once per
week. This normally runs fully automatically
Le GasLab Q2 doit effectuer un étalonnage opérationnel au moins une
fois par semaine, ce qui se fait normalement de manière entièrement auto-
matique.
8.4.1 Setting or changing automatic measurements
➪ 
GasLab Q2
GasLab Q2 operation panel and operation
User Manual (N )
(p. 236)