10
10.3 Inspection work in detail
10.3.1 Checking the battery charge
The battery of the device is mainly stressed when the voltage supply is permanently
switched off or is switched off and on again more frequently. When the voltage
supply is connected, the consumption of the battery can be neglected.
The battery must be replaced when the charge drops to less than 20%. In addition,
the battery loses energy due to aging and must be replaced with a new one after 10
years at the latest.Check the battery charge on the info display (device monitor).
NOTICE Discharged battery
AVIS Batterie déchargée
The device will no longer start correctly, Measurement data will be lost.
Check battery charge at least once a year.
L'appareil ne démarre plus correctement, les données de mesure sont per-
dues. Vérifiez la charge de la batterie au moins une fois par an.
Procedure:
On the device or the remote operation panel activate: Home- display – Info –
l
Device monitor
Read off the Battery state of charge and note the value in the maintenance
l
report.
Result evaluation and further procedure:
Charge > 50 % Battery can be used for another year.
l
Charge < 50 % Check battery more frequently
l
Charge < 20% Replace battery if necessary inform Honeywell for this purpose.
l
GasLab Q2
Maintenance, cleaning and repair
Fig. 10-1: Display Device Monitor
User Manual (N )
263 of 315