Isolated Patient Lead (Sink) Leakage-Isolation Test

Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter's function
selector to the LEAD-ISO. Select the ALL position on the lead selector. Depress the rocker switch to
ISO TEST to test lead isolation.
Line voltage is applied to the ECG leads during this test. To avoid possible electric shock
CAUTION
hazard, the system being tested must not be touched by patients, users or anyone while the ISO
TEST switch is depressed.
NOTE:
It is not necessary to test each lead individually or power condition combinations as required in
previous tests.
11
Chapter 10 - Care & Maintenance
V
7 S
M
IVID
ERVICE
ANUAL
10 - 19