Side Covers (With Bumpers) Replacement Procedure; Manpower; Tools; Preparations - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
Section 8-2

Side Covers (with Bumpers) Replacement Procedure

8-2-1

Manpower

One person, 5 minutes
8-2-2

Tools

Large coin and a large screwdriver
8-2-3

Preparations

Power off / Shutdown the system, see
8-2-4

Side Covers Removal Procedure

1.) Use a large coin to turn the two recessed fasteners at the bottom of the side cover a quarter turn to
the horizontal position.
2.) Take hold under the lower front side of cover and pull gently. At the same time support the rear side
of the cover with the other hand.
NOTE:
The two fasteners hold the left-side cover to the scanner at the bottom of the card cage cover.
The two fasteners hold the right-side cover to the scanner at the bottom of the AC power supply wall.
3.) When the cover is loose in front, pull the rear out and move upwards.
4.) Lift the side cover gently upwards until the three hooks at the top of the side cover are free from the
three retainer slots in the top cover as shown in
5.) To Remove Left side cover, repeat this procedure.
NOTE:
Figure 8-2
the left-side cover.
8 - 4
Section 8-2 - Side Covers (with Bumpers) Replacement Procedure
11
shows the right-side cover being removed. Repeat the same steps listed above to remove
Recessed 1/4 turn fasteners
Pull front side loose first
Figure 8-2 Removal of Side Covers
4-2-3 "Power Shut Down" on page
Figure
8-2.
Coin
V
7 S
M
IVID
ERVICE
ANUAL
4-7.

Advertisement

Table of Contents
loading