Set Up Connection to a DICOM Server in a Network

In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are
first saved on the local image buffer on the scanner. At the end of the examination the images are sent
to the DICOM server via a DICOM spooler and to the local database, depending on dataflows.
This scenario requires that the scanner is configured to be connected to DICOM servers as described
below.
3-9-14-1
Overview
To connect to the DICOM server, the following information has to be entered in the scanner:
The DICOM server IP address.
The DICOM server port number.
The DICOM server AE title (the server application's name).
Name of Device.
3-9-14-2
DICOM Server IP Address Setting on the Scanner
1.) Press CONFIG (F2) and log on as adm, see
2.) Select CONNECTIVITY (in the lower part of the window).
3.) Select the DATAFLOW tab.
4.) Select the arrow to the right for the Name field to list all dataflows in a pull-down menu, see
3-53.
5.) Select the dataflow you want to configure from the Dataflow Name pull-down list, see
3 - 58
Section 3-9 - Connectivity Setup - Software v7.x (BT'08), v6.x, v5.x and v4.x
11
Figure 3-53 Select Dataflow
4-2-4 "Log On to the System as 'ADM'" on page
V
7 S
M
IVID
ERVICE
ANUAL
4-12.
Figure
ARROW
SELECT DATAFLOW
FROM DATAFLOW NAME
PULL-DOWN MENU.
Figure
3-53.