Entering A Spot Enr Value; Changing The Default Tcold Value - Agilent Technologies NFA Series Quick Reference Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Step 1. Press the
Step 2. Enter an ENR value using the numeric keys and terminate it using the
NOTE
NOTE
Step 1. Press the
Step 2. Press the
Step 3. Press the
Making Basic Measurements
Entering ENR Data

Entering a Spot ENR Value

To enter a Spot ENR value:
key, then the
ENR
unit termination menu keys. The default value is 15.20 dB.
If the frequency you want to measure is not a listed ENR value, then you
need to interpolate the ENR list to an appropriate value.
To enable spot ENR mode to operate, press the
menu key.
ENR Mode(Spot)
Changing the Default T
When working in different temperature conditions you can change the
T
value to accommodate the condition.
cold
To change the T
value:
cold
key.
ENR
menu key changing it from the default
Tcold
.
Tcold(On)
User Tcold
Enter a T
value using the numeric keys and terminate it using the
cold
unit termination menu keys. The default T
The unit termination menu keys are in
(Fahrenheit).
F
28
menu key.
Spot ENR
value
cold
menu key
K
key, and select the
ENR
Tcold(Off)
value is 296.5K.
cold
(Kelvin),
(Celsius) or
C
to
Chapter 2