Reassembly - Siemens MAMMOMAT 3000 Repair Instructions

Advertisement

Procedure

Reassembly

1.
Remove the locking washer from the spare part wire kit with proper tool.
2.
Reassemble the spare part wires with handels.
NOTICE
3.
Secure the ring and the spring with the locking washer.
4.
Reassemble the biopsy unit.
NOTICE
Siemens-Elema AB
Solna, Sweden
All manuals and user guides at all-guides.com
Make sure that the track in the ring (see pos 2 in fig. 2)
fits to the hook of the spring.
Don´t forget to connect the grounding wire.
Register 5
SPB7-230.091.09
Rev. 02
Page 3 of 4
01.99
SPS-UD
2 - 3
2
MAMMOMAT 3000
Service